檢測(cè)認(rèn)證人脈交流通訊錄
ECG性能測(cè)試儀MECG2.0 多功能檢測(cè)儀及數(shù)據(jù)庫(kù)比對(duì)分析軟件播放器
- 用 途:
- 檢測(cè)心電圖機(jī)性能,符合YY0885,YY0782,YY1079,YY1139
- MECG 2.0心電數(shù)據(jù)庫(kù)播放器
全模擬訊號(hào)輸出的ECG數(shù)據(jù)庫(kù)播放器,協(xié)助您完成系統(tǒng)及算法驗(yàn)證,適用于研發(fā)及法規(guī)符合性測(cè)試。可直接加載及撥放各類(lèi)型心電數(shù)據(jù)庫(kù),符合國(guó)際及國(guó)內(nèi)最新版標(biāo)準(zhǔn):IEC 60601-2-25、IEC 60601-2-47、ANSI/AAMI EC57、YY0782、YY0885
產(chǎn)品詳情
產(chǎn)品簡(jiǎn)介
全模擬訊號(hào)輸出的ECG數(shù)據(jù)庫(kù)播放器,協(xié)助您完成系統(tǒng)及算法驗(yàn)證,適用于研發(fā)及法規(guī)符合性測(cè)試??芍苯蛹虞d及撥放各類(lèi)型心電數(shù)據(jù)庫(kù),符合國(guó)際及國(guó)內(nèi)新版標(biāo)準(zhǔn):IEC 60601-2-25、IEC 60601-2-47、ANSI/AAMI EC57、YY0782、YY0885
產(chǎn)品特點(diǎn)
可供心電診斷、可攜式心電設(shè)備仿真數(shù)據(jù)庫(kù)產(chǎn)生器
支持12 導(dǎo)聯(lián)心電圖機(jī),8 個(gè)信道同步產(chǎn)生數(shù)據(jù)庫(kù)訊號(hào)
內(nèi)建CAL、ANE 及Biological 波形
可直接加載及播放AHA、MIT、CU 數(shù)據(jù)庫(kù),無(wú)須格式轉(zhuǎn)換
較佳的輸出訊號(hào)準(zhǔn)確度,電壓輸出<±8μV
提供軟件開(kāi)發(fā)工具包(Software Development Kit) 讓用戶(hù)自行開(kāi)發(fā)全自動(dòng)化測(cè)試軟件
產(chǎn)品規(guī)格
Item Details / Reference Value
Output channels The 8 output channels are provided through
a network as specified in IEC 60601-2-51 to
provide signals to 10 lead electrodes; in the
device under test, this will be displayed as 12
leads. 8 outputs
10 lead
electrodes
12 leads
Voltage accuracy IEC 60601-2-51 specifies a limit of ±1%, but
does not provide a lower limit (all systems
must have a lower limit).
An inferred specification of 1% ±5μV is
derived from the device under test
specification in IEC 60601-2-51 of 5%
±25μV.
±1% for
values
greater or
equal to
500μV. ±5μV
for values
under 500μV.
Output voltage
The output voltage on most of the database
and ECG is +5mV~-5mV.
±5mV
Output noise
level 0-150Hz Output noise should not influence the test. A
value a 5μV is suitable for this requirement.
Can be verified by monitoring the signal in
the device under test using a “diagnostic”
filter setting. <5μV
Time accuracy IEC 60601-2-51 does not provide any limits.
An inferred limit from the device under test.
An inferred limit of ±1% is used (see 4.2).
The system’s design accuracy exceeds 0.1%
as a 100ppm crystal reference is used. ±1%
Sampling rate A maximum sampling rate of 1kHz matches
the sampling rates of ECG files. 1kHz
(8 channels)
Power supply Powered from the USB supply (5V 0.2A) N/A
Environment Intended for normal laboratory environment.
The selection of critical components such as
reference voltages, DAC, precision resistors
are known to be stable in the range shown. 15-30°C
10-95% RH
上海海悅電子科技有限公司
陳經(jīng)理
- [聯(lián)系時(shí)請(qǐng)說(shuō)明來(lái)自檢測(cè)通]
- 聯(lián)系方式:
-
請(qǐng)點(diǎn)擊查看電話(huà)
- 地址:
- 吳涇鎮(zhèn)和愛(ài)路40號(hào)
永登县|
勐海县|
禹城市|
维西|
民县|
孟连|
新竹市|
崇州市|
黄陵县|
上犹县|
牟定县|
沙河市|
左权县|
古田县|
财经|
赤壁市|
通州区|
兰考县|
田林县|
河东区|
广昌县|
无为县|
高雄县|
镇康县|
丰原市|
苍溪县|
北碚区|
马山县|
龙游县|
天门市|
台中市|
宣城市|
舞钢市|
汉寿县|
固安县|
定远县|
喀喇沁旗|
罗城|
四川省|
太和县|
桐庐县|